• Sp-8 Series Medical Automatic Single Double Channel Electric Infusion Syringe Pump Injection Pump
  • Sp-8 Series Medical Automatic Single Double Channel Electric Infusion Syringe Pump Injection Pump

Sp-8 Series Medical Automatic Single Double Channel Electric Infusion Syringe Pump Injection Pump

Certification: CE, ISO13485
Scale: Medium
Type: Infusion Pump
Medical Device Regulatory Type: Type 2
Weight: 2kg
Flow Rate Precision: ±3%
Customization:
Gold Member Since 2019

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company

Basic Info.

Model NO.
SP-8 Series
Transport Package
Carton
Specification
CE, ISO
Trademark
Audo
Origin
Jiangsu, China
HS Code
9018909090
Production Capacity
1000units Per Month

Product Description


SP-8A Single Channel Electric Infusion Syringe Pump Basic Parameter:
Dimensions(mm) 266(L)×141(W)×145(H)
Weight 2Kg
Power supply AC 100V~AC 242V,50/60HZ(±1 HZ)
Built-in batteries DC 12V 1800mAh
Battery Model JBT Ni-MH, rechargeable
Battery Continuous Use Time At least 4hrs(Fully charged new battery, under 5ml/h running)
Power consumption ≤25 VA
Infusion Flow Selection 0.1ml/h~1600ml/h, original value is 0.1ml/h
Bolus rate Syringe 50ml:1600ml/h
Syringe 30ml:900 ml/h
Syringe 20ml:600 ml/h
Syringe 10ml:400 ml/h
Infusion Range 0.1ml~9999.9ml, original value is 0.1ml
Flow Rate Precision ±3%
Occlusion Pressure High:110kPa±20kPa (825mmHg±150 mmHg)
Middle:80kPa±30kPa  (600mmHg±225mmHg)
Low:60kPa±20kPa  (450mmHg±150 mmHg)
The biggest output pressure < 975mmHg(130kPa)
Use Environment Temperature: 5ºC~43ºC
Relative Humidity: ≤90%RH (Under without condensation status)
Atmospheric Pressure: 860hPa~1060hPa
Storage Environment Temperature: -40ºC~+55ºC
Relative Humidity: ≤90%RH (Under without condensation status)
Atmospheric Pressure: 860hPa~1060hPa
Waterproof Level IPX4
Electric Shock Protection Class I, Type CF
Alarm Occlusion, nearly finish, finish, standby (no operation) low battery, battery exhausted etc.
Function Transfusion amount display, flow rate display, infused volume display

 SP-8B Double Channel Electric Infusion Syringe Pump Parameter:
Dimensions(mm) 271(L)×167(W)×287(H
Weight 3.6kg
Power supply AC 100V~AC 242V 50/60HZ(±1 HZ)
Built-in batteries DC 12V 1800mAh
Battery Model JBT Ni-MH, rechargeable
Battery Continuous Use Time At least 6hrs for single channel(Fully charged new battery, under 5ml/h running)
Power consumption ≤42 VA
Infusion Flow Selection 0.1ml/h~1600ml/h, original value is 0.1ml/h
Bolus rate Syringe 50ml:1600ml/h
Syringe 30ml:900 ml/h
Syringe 20ml:600 ml/h
Syringe 10ml:400 ml/h
Infusion Range 0.1ml~9999.9ml, original value is 0.1ml
Flow Rate Precision ±3%
Occlusion Pressure High:110kPa±20kPa (825mmHg±150 mmHg)
Middle:80kPa±30kPa  (600mmHg±225mmHg)
Low:60kPa±20kPa  (450mmHg±150 mmHg)
The biggest output pressure < 975mmHg(130kPa)
Use Environment Temperature: 5ºC~43ºC
Relative Humidity: ≤90%RH (Under without condensation status)
Atmospheric Pressure: 860hPa~1060hPa
Storage Environment Temperature: -40ºC~+55ºC
Relative Humidity: ≤90%RH (Under without condensation status)
Atmospheric Pressure: 860hPa~1060hPa
Waterproof Level IPX4
Electric Shock Protection Class I, Type CF
Alarm Occlusion, nearly finish, finish, standby (no operation) low battery, battery exhausted etc.
Function Transfusion amount display, flow rate display, infused volume display

 
Special Function
KVO (Keep Vein Open) After finish setting infusion amount, for prevent to form thrombus, keep to infusion at the lowest rate 0.1ml/h.
Power Switching Function When AC power is disconnected, automatically switch into battery power supply and give indication. When the AC power is connected, automatically switch into AC power working.
Repeat Alarm Function If the alarm real reason is not be resolved within 2 minutes after press "SILENCE" button, it will give alarm again to remind to resolve the problem.
Data Memory Function Able to save the latest setting alarm volume level, brand, occlusion pressure. level for use next time. Able to save for five years.
Battery Protection Alarm For protect battery and safety infusion, the battery power will be displayed as different levels in sound and light from nearly finish to exhausted
Move alarm When try to move or take the syringe away, it will give alarm.

 SP-8A/8B Syringe Pump's Advantages
1. Our syringe pump could use all brands through calibration (default setting is Terumo).
2. Auto-recognize the syringe size: 10, 20, 30, 50ml.
3. Wide flow rate range from 0.1-1600ml/h.
4. Available for adjust alarm volume (3 levels), occlusion pressure (3 levels).
5. Power Switching Function:When AC power is disconnected, automatically switch into battery power supply and give indication. When the AC power is connected, automatically switch into AC power working.
6. Repeat Alarm Function: If the alarm real reason is not be resolved within 1 minutes after press "SILENCE" button, it will give alarm again to remind to resolve the problem.
7. Battery Protection Alarm: For protect battery and safety infusion, the battery power will be displayed as different levels in sound and light from nearly finish to exhausted.
8. Double confirmation function when flow rate≥300ml/h, make safer and avoid misoperation.
9. The data can be cleared, make safer.
10. Lock the control panel during infusion, make safer.
11 For avoid misoperation, press "POWER" for 3 seconds to close the pump.
12. Real display the battery power in panel.
13. Different alarm light display in red and yellow color base different priority level.
14. RS232 port, able to connect with the computer.
15. KVO (Keep Vein Open) function. After finish setting infusion amount, for prevent to form thrombus, keep to infusion at the lowest rate 0.1ml/h.
16. Bolus (Quick Infuse) function.
17. Purge (Evacuate Bubble) function.
18. The bright & large LCD screen, easy look
19. The battery can be used continuously for 6 hours after full charged.
20. The chip in syringe pump is be processed by ourselves on SMT machine production line instead of by manual work, as weld every point by hand.
21. CPU control, stable function, high reliability.