Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Item Model |
SIP-10B | SIP-10C |
Dimensions(mm) | 354(L)×132(W)×199(H) | 354(L)×132(W)×268(H) |
Weight | 2.7kg | 3.6kg |
Battery Continuous Use Time | About 6 hours (fully charged new battery running with rate 5ml/h) |
About 5 hours (fully charged new battery running with rate 5ml/h) |
Rated Consumption | ≤25 VA | ≤35 VA |
Power Supply | AC 100V~AC 242V,50/60HZ |
Built-in Batteries | DC 12V 1800mAh, Ni-MH, rechargeable |
Infusion Flow Selection | 0.1ml/h~1200ml/h (1-150d/min), original value is 0.1ml/h (1d/min) |
Syringe Flow Selection | 0.1ml/h~1600ml/h, original value is 0.1ml/h |
Syringe Size | 5/6ml, 10/12ml, 20/25ml, 30/35ml,50/60ml |
Bolus Rate | Syringe 50ml:1600ml/h Syringe 30ml:900 ml/h Syringe 20ml:600 ml/h Syringe 10ml:400 ml/h Syringe 5ml: 200ml/h |
Infusion Range | 0.1ml~9999.9ml, original value is 0.1ml |
Infusion Precision | volume mode ±5%, drop mode ±2d |
Syringe Flow Rate Precision | ±3% |
KVO | 0.1~5ml/h |
Occlusion Pressure | High:110kPa±20kPa (825mmHg±150 mmHg) Middle:80kPa±30kPa(600mmHg±225mmHg) Low:60kPa±20kPa(450mmHg±150 mmHg) The biggest output pressure:< 975mmHg(130kPa) |
Use Environment | Temperature: 5ºC~43ºC Relative Humidity:≤90%RH (Under without condensation status) Atmospheric Pressure:860hPa~1060hPa |
Storage Environment | Temperature:-20ºC~+55ºC Relative Humidity:≤90%RH (Under without condensation status) Atmospheric Pressure:860hPa~1060hPa |
Waterproof Level | IPX4 |
Electric Shock Protection | Class I, Type CF |
Alarm: | occlusion, nearly finish, finish, standby (no operation), door open, low battery, battery exhausted etc. |
Function | power indicator, transfusion amount display, flow rate display, infused volume display etc. |