Specification
CE, , ISO
Trademark
Audo
Origin
Jiangsu
Production Capacity
1000units Per Month

Product Description

Product Feature

Temperature Control Mode: servo control

Instant Parameter Display: set temp, skin temp, heat power rate

Tilting Bed with Rotary Radiant Head

X-ray Cassette Tray under infant ben, examination light of high illuminance

Use anti-blast quartz infrared tube as radiant heating source

Specifications

Temperature Control Range: 20.0ºC-37.0ºC

Temperature sensor accuracy: ≤0.3ºC

Mattress surface temperature uniformity: ≤2.0ºC

Temperature rise time: ≤45mins

Heat output indicate: 0-100%

Radiant box horizon angle adjust range: 0-90°adjustable in step-less

Bed tilt adjustment: continues adjustable in ±10°

Mattress Tray Size: 68cm×54cm

Mattress Tray Max load: 10kg

Transfusion pole Max load: 2kg

Shelf Max Load: 2kg

Alarm

Sensor failure alarm

Over temperature alarm

Deviation alarm: ±1ºC

Power failure alarm

System failure alarm

Operation condition

Environment temperature: +18ºC- +30ºC

Environment relative humidity: 30%ºC-75%

Environment air velocity of flow: <0.3m/s

Irw-1000 Radiant Warmer for Infant, Medical Infant Radiant Warmer
 
 

STANDARD COMPOSITION

 

Servo Control Mode: air temperature

Instant Parameter Display: set temp, skin temp, heat power rate;

Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;

Tilting Bed with Rotary Radiant Head

X-ray Cassette Tray under infant bed, examination light of high illuminance;

Use anti-blast quartz infrared tube as radiant heating source.

 

 

 

TECHNICAL SPECIFICATION:

 

POWER

Model BNT-1000-1E.........................................................................................AC 110 / 120 V, 50Hz / 60Hz

Model BNT-1000-2E......................................................................................... AC 220 / 240V, 50Hz / 60Hz

Input Power..................................................................................................................................... ≤1000VA

TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION

Control mode....................................................................................................................................baby mode

Temperature control range.............................................................................................................25ºC~37ºC.

Skin temperature sensor measure range...........................................................................................20ºC~42ºC

Deviation between sensor measure temperature and control temperature.................................................≤0.5ºC

Skin Sensor Precision........................................................................................................................... ≤0.3ºC

Bed surface temperature Uniformity......................................................................................................... ≤1ºC

Temperature Rise Time ....................................................................................................................... ≤45min

Bed tiltable.............................................................................................................continues adjustable in ±10°

ALARM

Sensor failure alarm..................................................................alarm with audible and visual when sensor open

                                                                                                  circuit, short circuit or disconnect, cut off power.

Over temperature alarm................................................................alarm with audible and visual when displayed

                                                                                                                         temperature was approach 38ºC.

Deviation alarm................................................................after temperature equilibrium, alarm with audible and

                                                                                           visual when displayed temperature was 1ºC above or

                                                                                         below control temperature, cut off power if above 1ºC.

Power failure alarm................................................................ alarm with audible and visual when power failure

 

ENVIRONMENT TEMPERATURE

(It is a suggestion that please do not use outside of the allow environment.)

Operate temperature range................................................................................................................18ºC~30ºC

Store temperature range..............................................................................................................  -20ºC~+55ºC

ENVIRONMENT HUMIDITY

Operate humidity range........................................................................................................... 30%RH~75%RH

Store humidity range............................................................................................................................ ≤93%RH

ATMOSPHERIC PRESSURE

Transport and Store atmospheric pressure range......................................................................500hPa~1060hPa

Operate atmospheric pressure range.........................................................................................700hPa~1060hPa

STREAM VELOCITY

Ambit environment stream velocity.......................................................................................................... ≤0.3m/s