Customization: | Available |
---|---|
Customized: | Non-Customized |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Temperature Control Mode: servo control
Instant Parameter Display: set temp, skin temp, heat power rate
Tilting Bed with Rotary Radiant Head
X-ray Cassette Tray under infant ben, examination light of high illuminance
Use anti-blast quartz infrared tube as radiant heating source
Temperature Control Range: 20.0ºC-37.0ºC
Temperature sensor accuracy: ≤0.3ºC
Mattress surface temperature uniformity: ≤2.0ºC
Temperature rise time: ≤45mins
Heat output indicate: 0-100%
Radiant box horizon angle adjust range: 0-90°adjustable in step-less
Bed tilt adjustment: continues adjustable in ±10°
Mattress Tray Size: 68cm×54cm
Mattress Tray Max load: 10kg
Transfusion pole Max load: 2kg
Shelf Max Load: 2kg
Sensor failure alarm
Over temperature alarm
Deviation alarm: ±1ºC
Power failure alarm
System failure alarm
Environment temperature: +18ºC- +30ºC
Environment relative humidity: 30%ºC-75%
Environment air velocity of flow: <0.3m/s
STANDARD COMPOSITION |
|
Servo Control Mode: air temperature |
Instant Parameter Display: set temp, skin temp, heat power rate; |
Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override; |
Tilting Bed with Rotary Radiant Head |
X-ray Cassette Tray under infant bed, examination light of high illuminance; |
Use anti-blast quartz infrared tube as radiant heating source. |
|
|
|
TECHNICAL SPECIFICATION: |
|
POWER |
Model BNT-1000-1E.........................................................................................AC 110 / 120 V, 50Hz / 60Hz |
Model BNT-1000-2E......................................................................................... AC 220 / 240V, 50Hz / 60Hz |
Input Power..................................................................................................................................... ≤1000VA |
TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION |
Control mode....................................................................................................................................baby mode |
Temperature control range.............................................................................................................25ºC~37ºC. |
Skin temperature sensor measure range...........................................................................................20ºC~42ºC |
Deviation between sensor measure temperature and control temperature.................................................≤0.5ºC |
Skin Sensor Precision........................................................................................................................... ≤0.3ºC |
Bed surface temperature Uniformity......................................................................................................... ≤1ºC |
Temperature Rise Time ....................................................................................................................... ≤45min |
Bed tiltable.............................................................................................................continues adjustable in ±10° |
ALARM |
Sensor failure alarm..................................................................alarm with audible and visual when sensor open |
circuit, short circuit or disconnect, cut off power. |
Over temperature alarm................................................................alarm with audible and visual when displayed |
temperature was approach 38ºC. |
Deviation alarm................................................................after temperature equilibrium, alarm with audible and |
visual when displayed temperature was 1ºC above or |
below control temperature, cut off power if above 1ºC. |
Power failure alarm................................................................ alarm with audible and visual when power failure |
|
ENVIRONMENT TEMPERATURE |
(It is a suggestion that please do not use outside of the allow environment.) |
Operate temperature range................................................................................................................18ºC~30ºC |
Store temperature range.............................................................................................................. -20ºC~+55ºC |
ENVIRONMENT HUMIDITY |
Operate humidity range........................................................................................................... 30%RH~75%RH |
Store humidity range............................................................................................................................ ≤93%RH |
ATMOSPHERIC PRESSURE |
Transport and Store atmospheric pressure range......................................................................500hPa~1060hPa |
Operate atmospheric pressure range.........................................................................................700hPa~1060hPa |
STREAM VELOCITY |
Ambit environment stream velocity.......................................................................................................... ≤0.3m/s |