Customization: | Available |
---|---|
Type: | Leep Surgery System |
Kind: | Surgery |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications
LP-1D LEEP Surgery System was composed of LP-1 Trolley smoke vacuum and BC-50D High Frequency Electrosurgical Unit
1. BC-50D High Frequency Electrosurgical Unit
1.1. Classification: Class I Type CF
1.2. Main Frequency: 625kHz±5%
1.3. The mode of operation: CONTINUOUS OPERATION WITH INTERMITTENT LOADING, 10s/30s
1.4. Output power and peak voltage:
Spec. Mode |
Output Power |
|
Power Range |
Resistance |
|
Monopolar Pure cutting |
5~350W |
500Ω |
Monopolar Blend 1 |
5~250W |
500Ω |
Monopolar Blend 2 |
5~200W |
500Ω |
Monopolar Contact Coagulation |
5~200W |
500Ω |
Bipolar Standard Coagulation |
2~70W |
150Ω |
1.5. Automatic Storage: The working mode and output power set during operation can be stored in the unit.
1.6. High Frequency Leakage Current
The neutral electrode is isolated from the earth during the high frequency operation, the High Frequency Leakage Current <150mA.
1.7. Output Channels: Monopolar: 1 channel, Bipolar: 1 channel.
1.8. Output Setting and Indicating: The output is set by key pressing in different area according to different working mode, it is displayed digitally.
1.9. Output Indicating
Visual Indicator: Monopolar Cutting - Yellow / Coag - Blue / Bipo - Blue
Audio Indicator: When the power is output, there is an audio indicator.
1.10. Error Indicating
Visual Indicator: The red indicator indicating the error and danger detected by self-detecting or monitoring.
Audio Indicator: When alarming, there is an pulse audio indicating.
1.11. Output Control
Monopolar: Hand Control, Foot Control.
Bipolar: Foot Control, Automatic Control.
2. LP-1 Trolley smoke vacuum
2.1. Maximum air flow: 100L/min
2.2. Ventilate Strength: 10 - 100% flow volume: 10L/min-100L/min
2.3. Working Noise: < 60db
2.4. Particle Remove Rate: 99.999% @0.1um
2.5. Bad Smell Removal Rate: 97% @1year
3. After the trolley integration
3.1. Power supply: ~220V,50Hz
3.2. Max Input Power: 1200VA
3.3. Dimensions: :460mm×440mm×800mm
3.4. Net Weight: 22kg
3.5. Use condition: environment temperature +5ºC~+40ºC relative humidity≤85%
3.6. Transportation and storage condition: environment temperature -40ºC~+50ºC relative humidity≤90%
Contact us: