Customization: | Available |
---|---|
Type: | Electric Syringe Pump |
Kind: | Surgery |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product name
|
Medical Syringe Pump Electric Injection Pump Hospital Use Syringe Pump |
Infusion rate
|
5 ml syringe : 0.1-150ml/h
10ml syringe: 0.1-300ml/h
20ml syringe: 1.0-600ml/h
30ml syringe: 0.1-900ml/h
50ml syringe: 0.1-1500ml/h
|
Syringe accuracy
|
±2%(After syringe correct calibration)
|
Applicable syringes
|
Various brands of sterile syringes(Meet GB15810-2001 standard): 5ml, 10ml, 20ml, 30ml, 50ml/60ml
|
Volume to be infused
|
0-1000(ml)
|
Volume infused
|
0-9999.9(ml)
|
KVO rate
|
0.1-2ml/h adjustable; default value 0.1ml/h
|
BOLUS(ML/H)
|
5 ml syringe : 100-150ml/h
10ml syringe: 100-300ml/h
20ml syringe: 100-600ml/h
30ml syringe: 100-900ml/h
50ml syringe: 100-1500ml/h
|
Infusion mode
|
rate mode
time mode
Body Weight mode
|
History record
|
1600 records
|
Computer interface
|
RS232 interface
|
Alarm
|
Syringe disengaged, Near compoletion, Infusion completion, Empty, Occlusion, Low Battery, handle off etc
|
Power supply
|
DC:12±1.2V
AC:100~240V, 50/60HZ
|
Nurse call interface
|
|
Battery
|
Rechargeable Li Polymer battery, 7.4V 1900mAh
|
Battery charging time
|
10 hours with power on, 3 hours with power off
|
Battery operating time
|
>6 hours at 5ml/h
|
Power consumption
|
25 VA
|
Classification
|
Class l, type CF
|
Dimension
|
300x125x130mm
|
Net Weight
|
1.8kg
|
Waterproof level
|
IPx3
|
Occlusion pressure
|
(40.0-60.0)kpa 3 adjustable occlusion pressure : low middle, high
|
Operating conditions
|
Temperature :5°C-40 °C, relative humidity :10-95%
Atmosphere pressure:86-106kpa
|
Transport conditions
|
Tempearture:-20ºC-60ºC,relative humidity:10-95%
Atmosphere pressure:50-106kPa
|
Storage conditions
|
Temperature :5- 20 °C-45 °C, relative humidity :10-95%